Exfo CableSHARK P3 VF/DSL Cable Qualifier Manual de usuario Pagina 223

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 321
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 222
CableSHARK P3 User Guide
211
5. For the grounded system, the CableSHARK switches the loop current path with the 435 ohm load
(currently off hook), from tip-ring to tip-ground. The CableSHARK measure both the ring-ground
voltage and the loop current.
The Ground Resistance measurement sequence is based on the following assumptions:
1. The ground lead of the instrument is connected reliably to the ground of the CO
2. The standard CO voltage is nominally -48 VDC. The twisted pair is connected to the CO with tip as
positive and ring as negative.
3. The grounded constant current source limits the current on the –48 V side only.
Vista de pagina 222
1 2 ... 218 219 220 221 222 223 224 225 226 227 228 ... 320 321

Comentarios a estos manuales

Sin comentarios