Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 600

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 599
Referencia de instrucciones SCPI
594 FTB-5240S/S-P/BP
Product-Specific Commands—Description
The <Condition> response represents the
current operation condition of the instrument.
The meaning of the response depends on the
value returned for bit <n>.
bit <8>: When the returned value is 1, the
instrument is performing a measurement (trigger
system INITiated).
bit <9>: When the returned value is 1, the
instrument is performing an offset nulling and/or
a wavelength referencing
(CALibration:ZERO:AUTO?).
Example(s) STAT? Must return READY
CAL:ZERO:AUTO ONCE
STAT:OPER:BIT9:COND? Keep resending this
query as long as the operation is not complete
(returned value is not 0).
See Also :CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus?
:STATus:OPERation:BIT<8|9>:
CONDition?
Vista de pagina 599
1 2 ... 595 596 597 598 599 600 601 602 603 604 605 ... 658 659

Comentarios a estos manuales

Sin comentarios