Referencia de instrucciones SCPI
OSA 563
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:STATe?
Description This query indicates if acquired trace averaging
has been enabled or not.
At *RST, averaging is set to off (disabled).
Syntax :SENSe[1..n]:AVERage:STATe?
Parameter(s) None
Response Syntax <State>
Response(s) State:
The response data syntax for <State> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <State> response corresponds to the
activation state of trace averaging.
0: trace averaging is disabled.
1: trace averaging is enabled.
Example(s) SENS:AVER:STAT ON
SENS:AVER:STAT? Returns 1 (averaging is
enabled)
See Also :SENSe[1..n]:AVERage:COUNt
:SENSe[1..n]:AVERage:TYPE
:SENSe[1..n]:AVERage:STATe