Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 579

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 578
Referencia de instrucciones SCPI
OSA 573
Product-Specific Commands—Description
:SENSe[1..n]:FREQuency:STARt?
Description This query returns a value indicating either the
current or the minimum/maximum instrument
sweep start frequency.
At *RST, this value is set to 190.9506 THz.
Syntax :SENSe[1..n]:FREQuency:STARt?[<wsp>MAXim
um|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Start>
Vista de pagina 578
1 2 ... 574 575 576 577 578 579 580 581 582 583 584 ... 658 659

Comentarios a estos manuales

Sin comentarios