Referencia de instrucciones SCPI
434 FTB-5240S/S-P/BP
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Width> parameter.
The <Width> parameter corresponds to a valid
channel width in hertz.
The
CALCulate[1..n][:WDM]:CHANnel:WIDTh:FREQ
uency? MIN and
CALCulate[1..n][:WDM]:CHANnel:WIDTh:FREQ
uency? MAX queries can be used to determine
valid channel frequency width.
Example(s) CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ
CALC:WDM:CHAN:SEL "ITU_22"
CALC:WDM:CHAN:WIDT:FREQ 200.0 GHZ
CALC:WDM:CHAN:WIDT:FREQ? Returns
2.000000E+011
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO:WIDTh
:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:WIDTh[:WAV
elength]
:CALCulate[1..n][:WDM]:CHANnel:WIDTh:FREQ
uency?
:CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:CHANnel:
WIDTh:FREQuency