Referencia de instrucciones SCPI
OSA 425
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
CENTer:FREQuency
Description This command sets the nominal center
frequency of the selected WDM analysis channel.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ
uency<wsp><Center[<wsp>HZ]>|MAXimum
|MINimum|DEFault
Parameter(s) Center:
The program data syntax for <Center> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is HZ. The <Center>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.