Referencia de instrucciones SCPI
OSA 501
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:DATA:CHANnel:
NOISe:AUTO?
Description This query indicates if the selected WDM
channel result was computed using i-InBand
auto noise measurement.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:DATA:CHANnel:NOISe:
AUTO?
Parameter(s) None
Response Syntax <Auto>
Response(s) Auto:
The response data syntax for <Auto> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Auto> response corresponds to the state
of the auto noise measurement.
0: auto noise measurement is disabled.
1: auto noise measurement is enabled.