Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 385

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 384
Referencia de instrucciones SCPI
OSA 379
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:AUTO
Description This command controls activation of i-InBand
noise measurement for WDM analysis default
channel.
At *RST, auto noise is set to off (disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of auto noise measurement.
0 or OFF: disable default channel auto noise.
1 or ON: enable default channel auto noise.
Vista de pagina 384
1 2 ... 380 381 382 383 384 385 386 387 388 389 390 ... 658 659

Comentarios a estos manuales

Sin comentarios