Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 591

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 590
Referencia de instrucciones SCPI
OSA 585
Product-Specific Commands—Description
:SENSe[1..n][:WAVelength]:STARt?
Description This query returns a value indicating either the
current or the minimum/maximum instrument
sweep start wavelength.
At *RST, this value is set to 1525.0 nm.
Syntax :SENSe[1..n][:WAVelength]:STARt?[<wsp>MAXi
mum|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Start>
Vista de pagina 590
1 2 ... 586 587 588 589 590 591 592 593 594 595 596 ... 658 659

Comentarios a estos manuales

Sin comentarios