Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 598

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 597
Referencia de instrucciones SCPI
592 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:STATus?
Description This query returns a value indicating the global
status of the instrument.
This command has no associated *RST
condition.
Syntax :STATus?
Parameter(s) None
Response Syntax <Status>
Response(s) Status:
The response data syntax for <Status> is defined
as a <CHARACTER RESPONSE DATA> element.
The <Status> response represents the
instrument state, where:
UNINITIALIZED means the instrument has not
been initialized yet.
INITINPROGRESS means the instruments
initialization is in progress.
READY means the instrument is ready.
BUSY means the instrument is busy.
DISCONNECTED means the instrument is
disconnected.
DEFECTIVE means the instrument is defective.
Example(s) STAT?
See Also :CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus:OPERation:BIT<8|9>:CONDition?
Vista de pagina 597
1 2 ... 593 594 595 596 597 598 599 600 601 602 603 ... 658 659

Comentarios a estos manuales

Sin comentarios