Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 469

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 468
Referencia de instrucciones SCPI
OSA 463
Product-Specific Commands—Description
Example(s) CALC:WDM:CHAN:DEF "C_001", 1290.000 NM
CALC:WDM:CHAN:SEL "C_001"
CALC:WDM:CHAN:NOIS:AUTO OFF
CALC:WDM:CHAN:NOIS:TYPE INBN
CALC:WDM:CHAN:NOIS:TYPE? Returns
INBANDNARROWFILTER
Notes INBand and INBandNarrowfilter noise types are
available only if software option "InB" is active.
INBand and INBandNarrowfilter noise types are
computed only if analysed trace was acquired
using PMMH averaging type.
If auto noise measurement is active, specific
noise type setting has no effect.
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:AUTO
:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE?
:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan
ce:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan
ce[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT
h:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT
h[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:SELect
:SENSe[1..n]:AVERage:TYPE
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:TYPE
Vista de pagina 468
1 2 ... 464 465 466 467 468 469 470 471 472 473 474 ... 658 659

Comentarios a estos manuales

Sin comentarios