Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 380

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 379
Referencia de instrucciones SCPI
374 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO
Description This command controls the state of the WDM
analysis default channel (enabled or diabled).
At *RST, the state of the default channel is set to
on (enabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO<wsp>
<Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of the default channel.
0 or OFF: disable default channel.
1 or ON: enable default channel.
Example(s) CALC:WDM:CHAN:AUTO ON
CALC:WDM:CHAN:AUTO? Returns: 1 (default
channel is enabled)
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO?
:CALCulate[1..n][:WDM]:CHANnel:CATalog?
:CALCulate[1..n][:WDM]:CHANnel[:DEFine]
:CALCulate[1..n][:WDM]:DATA:CHANnel:CATalo
g?