Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 441

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 659
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 440
Referencia de instrucciones SCPI
OSA 435
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
WIDTh:FREQuency?
Description This query returns the frequency width of the
selected WDM analysis channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to 50.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:WIDTh:FREQ
uency?[<wsp>MAXimum|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Width>
Vista de pagina 440