Remote Control Commands
88 PM-1100
Product-Specific Commands—Description
SENSitivity:AVERage[:STATe]
Description This command activates or deactivates data averaging.
Syntax SENS:AVER[:STAT]<space><boolean>
Parameters The < boolean > parameter can be
“0” averaging is disabled
“1” averaging is enabled
Example SENS:AVER:STAT 1
SENSitivity:AVERage:STATe?
Description This query returns a value indicating whether data averaging is
enabled or disabled.
Syntax SENS:AVER:STAT?
Response “0” averaging is disabled
“1” averaging is enabled
Example SENS:AVER:STAT?
SENSitivity:CORRection:COLLect:ZERO
Description This command performs an offset nulling measurement.
Syntax SENS:CORR:COLL:ZERO
Example SENS:CORR:COLL:ZERO