Exfo PM-1100 Optical Power Meter Manual de usuario Pagina 96

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 111
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 95
Remote Control Commands
86 PM-1100
Product-Specific Commands—Description
MMEMory:ACQuisition:SAMPles
Description This command changes the samples parameter for the
acquisition setup.
Syntax MMEM:ACQ:SAMP<space><samples>
Parameters The <samples> parameter is the number of samples to be set
for the acquisition setup. If an invalid parameter is entered, the
closest valid parameter will be entered instead.
Note The duration of the acquisition has a direct effect on the values
that can be set for the number of samples.
D x R = S
where
D = duration parameter set with the MMEM:ACQ:DURA
command (in seconds)
R = any of the possible sampling rates in sec
-1
(get the
complete list with the SENS:FREQ:CAT? query)
S = valid number of samples (MUST be a whole number)
Example MMEM:ACQ:SAMP 200
MMEMory:ACQuisition:SAMPles?
Description This command returns the current number of samples for the
acquisition setup.
Syntax MMEM:ACQ:SAMP?
Response Current number of samples in the “1” format
Example MMEM:ACQ:SAMP?
Vista de pagina 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 110 111

Comentarios a estos manuales

Sin comentarios