Remote Control Commands
90 PM-1100
Product-Specific Commands—Description
SENSitivity:FREQuency:CATalog?
Description This query returns a list of available sampling rates.
Syntax SENS:FREQ:CAT?
Response List of available sampling rates in the
“40.0;20.0;10.0;5.0;1.0;0.5;0.1” format
Example SENS:FREQ:CAT?
SENSitivity:POWer:REFerence?
Description This query returns the reference power for the current
wavelength.
Syntax SENS:POW:REF?
Response The current reference value in dBm, which can be any value
within the power range of the power meter in the “99.999 dBm”
format.
Example SENS:POW:REF?
SENSitivity:POWer:REFerence:DISPlay
Description This command performs a new reference measurement for the
current wavelength and changes the display to read relative
power (dB units).
Syntax SENS:POW:REF:DISP
Example SENS:POW:REF:DISP
Comentarios a estos manuales