Remote Control Commands
Optical Power Meter 83
Product-Specific Commands—Description
MMEMory:ACQuisition:DATA?
Description This command returns the number of data saved by the last
acquisition.
Syntax MMEM:ACQ:DATA?
Response Number of acquired data saved in memory locations in the
“600” format.
Example MMEM:ACQ:DATA?
MMEMory:ACQuisition:DATA:RECAll:UNIT?
Description This query returns the measurement units corresponding to the
last value fetched with the MMEM:ACQ:DATA:RECA:VALUE?
query.
Syntax MMEM:ACQ:DATA:RECA:UNIT?
Response The measurement units in the format
“0” for dBm
“1” for watt
“2” for dB
“3” for delta watt
Example MMEM:ACQ:DATA:RECA:UNIT?