Optical Power Meter 95
C SCPI-Based Errors
Error
Number
Description Probable Cause
-100 “Command error.” An error occurred while validating a
command.
-101 “Undefined header.” Unknown command.
-102 “Missing parameter.” A command parameter is missing.
-103 “Parameter not allowed.” An extra parameter is present.
-104 “Data type error.” Invalid parameter format.
-200 “Execution error.” An error occurred while executing a
command.
-300 “Device dependent error.” A command has taken longer than
expected to complete execution.
-300 “Light on detector.” Light reached the detector during
the offset nulling.
-300 “Invalid wavelength.” The entered wavelength is out of
range.
-300 “Parameter out of range.” The command parameter is out of
range.
-300 “No stored data.” No data is stored in the specified
location.
-300 “Program running.” An acquisition is already in progress.
-300 “Program duration not
initialized.”
No duration parameter is currently
defined for the program.
-300 “No samples available.” No sample parameter is currently
defined for the program.
-300 “No data.” No data to recall.
-300 “Invalid unit.” The requested measurement units
are invalid.
-300 “Invalid resolution.” The requested resolution is invalid.