Remote Control Commands
78 PM-1100
Product-Specific Commands—Description
DISPlay:DIMMer?
Description This query returns the intensity of the PM-1100 display (high,
low, or off)
Syntax DISP:DIMM?
Response “HI” if the intensity is high
“LO” if the intensity is low
“OFF” if the display is off
Example DISP:DIMM?
FETCh[:SCALar]:POWer:DC?
Description This query returns the last stored value.
Syntax FETC(0..26)[:SCAL]:POW:DC?
Response A power measurement in the “±999.999E±99” exponential
format in the selected units. The number of digits after the
decimal point depends on the selected resolution.
Example FETC3:SCAL:POW:DC?
See also INIT:CONT, INIT:CONT?, INIT:IMM, and READ:SCAL:POW:DC?