Remote Control Commands
Optical Power Meter 77
Product-Specific Commands—Description
Product-Specific Commands—Description
ABORt
Description This command stops any measurement in progress.
Syntax ABOR(0..26)
Example ABOR3
DISPlay:DIMMer
Description This command is used to adjust the intensity of the PM-1100
display (high or low) or to turn off the display without turning off
the unit.
Syntax DISP:DIMM<space><data>
Parameters The <data> parameter can be “HI”, “LO”, or “OFF”.
Example DISP:DIMM OFF
Note When the display is turned off by this command, any key
pressed on the PM-1100 keypad will return the display to high
intensity. To prevent this, lock the PM-1100 keypad with the
*LOK command.
See also DISP:DIMM?, *LOK, and *LOK?