Exfo AXS-200/850 Ethernet Test Set Manual de usuario Pagina 117

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 262
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 116
BERT
Ethernet Test Set 107
Summary Results
³ BER Threshold: Enter the threshold amount as follows (if the bit error
results exceed this value, the test will be declared as failed):
³ If the Pass/Fail Verdict is Bit Error Count, the accepted threshold
range is 6 digits with a default value of 0.
³ If the Pass/Fail Verdict is Bit Error Rate, the accepted threshold
range is 1.0E-14 to 1.0E0 with a default value of 1.0E-2.
³ The Pass/Fail Verdict is displayed on the screen with the help of
the following symbols:
= PASS
= FAIL
³ Reset: Press this button at any time to clear the results and statistics.
This does not interrupt the monitoring process nor cause errors on the
transmitted stream while a test is running. Use after an injection of bit
errors to clear results.
³ Bit Error Count: Displays the sum of mismatch ‘0’ and mismatch ‘1’
errors.
³ Bit Error Rate: Displays the bit error rate with a range of 1.0E-14 to
1.0E0.
³ Report: Select the Report button to generate the report file of the
results and statistics. Refer to Report Generation on page 79 to
generate and save a report file.
Vista de pagina 116
1 2 ... 112 113 114 115 116 117 118 119 120 121 122 ... 261 262