Exfo FTB-800 Series for FTB-1 Manual de usuario Pagina 509

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 621
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 508
Maintenance
NetBlazer 499
Recalibrating the Unit
Recalibrating the Unit
EXFO manufacturing and service center calibrations are based on the
ISO/IEC 17025 standard (General Requirements for the Competence of
Testing and Calibration Laboratories). This standard states that calibration
documents must not contain a calibration interval and that the user is
responsible for determining the re-calibration date according to the actual
use of the instrument.
The validity of specifications depends on operating conditions. For
example, the calibration validity period can be longer or shorter depending
on the intensity of use, environmental conditions and unit maintenance, as
well as the specific requirements for your application. All of these elements
must be taken into consideration when determining the appropriate
calibration interval of this particular EXFO unit.
Under normal use, the recommended interval for your FTB-700G/800
Series is: 2 years.
For newly delivered units, EXFO has determined that the storage of this
product for up to six months between calibration and shipment does not
affect its performance (EXFO Policy PL-03).
Vista de pagina 508
1 2 ... 504 505 506 507 508 509 510 511 512 513 514 ... 620 621

Comentarios a estos manuales

Sin comentarios