Test Results
NetBlazer 349
Summary (EtherSAM)
Service Configuration/Performance Test Buttons
Tap Service Configuration Test or Service Performance Test button to
view the result summary of the corresponding test.
Service indicates the service’s number and name.
Direction, available with Dual Test Set, indicates results from local to
remote (L -> R) and remote to local (R -> L).
Service Performance Test column displays the pass/fail verdict icon
indicating if the service complies to the configured SLA parameters.
Service Configuration Test column displays the pass/fail verdict icon
indicating if the service complies or not to the configured SLA
parameters.
Committed
Frame Loss indicates the percentage of frames that are lost. The
reported value is the maximum percentage of Frame Loss from all
burst sequences and ramp steps excluding the CIR+EIR, EBS, and
Traffic Policing steps.
Max. Jitter (ms) indicates the maximum measured delay variation.
Max Latency (ms) indicates the maximum measured round-trip
latency (delay). For Dual Test Set the local to remote and remote to
local values are reported for One-Way Latency Measurement Mode
while a single round-trip value is reported for Round-Trip Latency
Measurement Mode (see Global Options on page 123).
Avg RX Rate, for Service Performance Test, indicates the measured
average utilization throughput.
Excess
Max RX Rate, for Service Configuration Test, indicates the measured
maximum utilization throughput.