SCPI Command Reference
OSA 603
Examples on Using the SCPI Commands
Cancelling the Current Acquisition Sequence
Configuring the Analysis Setup for the Next
Acquisition Sequence (WDM)
Command Sequence Comments
SENS:AVER:STAT ON
SENS:AVER:TYPE SCAL
SENS:AVER:COUN 500
TRIG:SEQ:SOUR IMM
STAT?
INIT:IMM Start averaging acquisition.
ABOR Stop acquisition.
Command Sequence Comments
CALC:WDM:STAT ON Activate WDM analysis.
TRAC:FEED:CONT "TRC1", ALW Set trace data refresh mode to ALWays. When a
new sensed trace is available, it is
automatically transferred in the WDM calculate
block for analysis.
<Add commands to set global
parameters>
See Modifying Global Analysis Parameters
(WDM) on page 606.
<Add commands to configure
channel list>
See Creating a Channel List Based on the
Default Channel (WDM) on page 604 and
Creating a Channel List Based on Specific
Channels (WDM) on page 605.