Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual de usuario Pagina 383

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 638
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 382
SCPI Command Reference
OSA 377
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:WIDTh[WAVelength]?
Description This query returns the wavelength width of the
noise measurement region of the WDM analysis
default channel.
At *RST, the width of the default channel noise
measurement region is set to 100.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh[WAVelength]?[<wsp>MAXimum|MINi
mum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Width>
Vista de pagina 382
1 2 ... 378 379 380 381 382 383 384 385 386 387 388 ... 637 638

Comentarios a estos manuales

Sin comentarios