Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual de usuario Pagina 428

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 638
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 427
SCPI Command Reference
422 FTB-5240S/S-P/BP
Product-Specific Commands—Description
Example(s) CALC:WDM:CHAN:DEF "C_001",192.1750 THZ
CALC:WDM:CHAN:SEL "C_001"
CALC:WDM:CHAN:NOIS:AUTO ON
CALC:WDM:CHAN:NOIS:AUTO? Returns 1 (auto
noise enabled)
Notes Auto noise is available only if software option
"InB" is active.
Auto noise is computed only if analysed trace
was acquired using PMMH averaging type.
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO
:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:AUTO?
:CALCulate[1..n][:WDM]:CHANnel:SELect
:SENSe[1..n]:AVERage:TYPE
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:AUTO