Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual de usuario Pagina 515

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 638
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 514
SCPI Command Reference
OSA 509
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]:AUTO?
Description This query indicates if WDM analysis OSNR
calculation using auto resolution bandwidth has
been enabled or not for all channels.
At *RST, auto resolution bandwidth is set to off
(disabled).
Syntax :CALCulate[1..n][:WDM]:OSNR:BANDwidth|BWI
Dth[:RESolution]:AUTO?
Parameter(s) None
Response Syntax <Auto>
Response(s) Auto:
The response data syntax for <Auto> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Auto> response corresponds to the state
of the auto resolution bandwidth for OSNR
calculation.
0: custom resolution bandwidth value is used.
1: instruments resolution bandwidth is used.
Example(s) CALC:OSNR:BAND:RES:AUTO ON
CALC:OSNR:BAND:RES:AUTO? Returns 1
(instrument's RBW enabled)
See Also :CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDw
idth[:RESolution]
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDw
idth[:RESolution]:AUTO
Vista de pagina 514
1 2 ... 510 511 512 513 514 515 516 517 518 519 520 ... 637 638

Comentarios a estos manuales

Sin comentarios