Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual de usuario Pagina 457

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 638
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 456
SCPI Command Reference
OSA 451
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
SIGnalPower:TYPE
Description This command selects signal power
measurement type for WDM analysis of the
selected channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to IPOWer (integrated power).
Syntax :CALCulate[1..n][:WDM]:CHANnel:SIGnalPower:
TYPE<wsp>IPOWer|PPOWer|TPOWer
Parameter(s) Type:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
IPOWer|PPOWer|TPOWer.
The parameter corresponds to the newly
selected signal power type.
IPOWer: selects integrated signal power type.
PPOWer: selects peak signal power type.
TPOWer: selects channel total power type.
Vista de pagina 456
1 2 ... 452 453 454 455 456 457 458 459 460 461 462 ... 637 638

Comentarios a estos manuales

Sin comentarios