Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual de usuario Pagina 372

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 638
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 371
SCPI Command Reference
366 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:DISTance:[WAVelength]
Description This command sets the wavelength distance
from peak to center of noise region for noise
measurement of the WDM analysis default
channel.
At *RST, default channel noise measurement
distance is set to 100.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:[WAVelength]<wsp><Distance[<ws
p>M]>|MAXimum|MINimum|DEFault
Parameter(s) Distance:
The program data syntax for <Distance> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is M. The
<Distance> special forms MINimum, MAXimum
and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Vista de pagina 371
1 2 ... 367 368 369 370 371 372 373 374 375 376 377 ... 637 638

Comentarios a estos manuales

Sin comentarios