SCPI Command Reference
OSA 389
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
WIDTh:FREQuency?
Description This query returns the frequency width of the
WDM analysis default channel.
At *RST, default channel width is set to 50.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:WIDTh
:FREQuency?[<wsp>MAXimum|MINimum|DEF
ault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.