Setting Up the Instrument in EDFA Mode
OSA 227
Setting Up EDFA Analysis Parameters
Name prefix: adds prefix to the channel names.
Starting Value: sets the increment starting value for the channel
name in the channel list.
Increment value: sets the increment value for the channel name in
the channel list.
5. Press OK to return to the Channels window, which now lists the added
channels.
Note: When new channels are added, the Use Default thresholds will be applied
to the channel parameters.
Note: A warning message will be displayed if any channels are overlapping, but
the analysis can still be performed on overlapping channels. If any
duplicate channels are added, a confirmation message will be displayed to
overwrite the existing channels with the duplicate channels.
6. Press OK to save the changes and close the window, or press Cancel to
exit without saving.
Note: The application displays a message if more than 1000 channels are added.
You ca n exi t t he Analysis Setup window only after deleting the extra
channels from the channel list. You can delete the channels manually as
required.