Setting Up the Instrument in WDM Mode
70 FTB-5230S/-OCA
Setting Up WDM Analysis Parameters
4. Under Global analysis parameters, define the following parameters
as needed:
Peak detection level (dBm): indicates the minimum power level
from where the peak can be considered as a signal.
RBW for OSNR (nm): indicates the resolution bandwidth selected
for the OSNR calculation. This parameter is generally set to 0.1 nm
to allow for a common basis of comparison between different
OSA/OCAs having different effective resolutions. The instrument’s
RBW value is written below the graph. This parameter does not
actually have an effect on the acquisition, but is only a
normalization factor used to provide the OSNR value in a
standardized manner.