Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 224

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 251
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 223
Maintenance
218 FTB-5230S/-OCA
Cleaning EUI Connectors
To clean EUI connectors:
1. Remove the EUI from the instrument to expose the connector
baseplate and ferrule.
2. Moisten a 2.5 mm cleaning tip with one drop of isopropyl alcohol
(alcohol may leave traces if used abundantly).
3. Slowly insert the cleaning tip into the EUI adapter until it comes out on
the other side (a slow clockwise rotating movement may help).
4. Gently turn the cleaning tip one full turn, then continue to turn as you
withdraw it.
5. Repeat steps 3 to 4 with a dry cleaning tip.
Note: Make sure you don’t touch the soft end of the cleaning tip.
WARNING
Looking into the optical connector while the light source is active
WILL result in permanent eye damage. EXFO strongly recommends
to TURN OFF the unit before proceeding with the cleaning
procedure.
Push
Turn
Pull
3
4
5
Vista de pagina 223