Exfo FTB-5230S Optical Spectrum Analyzer for FTB-500 Manual de usuario

Busca en linea o descarga Manual de usuario para Instrumentos de medición Exfo FTB-5230S Optical Spectrum Analyzer for FTB-500. EXFO FTB-5230S Optical Spectrum Analyzer for FTB-500 User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 251
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente

Indice de contenidos

Pagina 1 - FTB-5230S/-OCA

User GuideOptical Spectrum and Channel Analyzersfor FTB-500FTB-5230S/-OCA

Pagina 2

Introducing the FTB-5230S/-OCA Optical Spectrum and Channel Analyzers4 FTB-5230S/-OCAConventionsConventionsBefore using the product described in this

Pagina 3 - Contents

Setting Up the Instrument in Drift Mode94 FTB-5230S/-OCADefining Preferences2. Select the General tab.3. Define the general parameters as needed.4. Pr

Pagina 4

Setting Up the Instrument in Drift ModeOSA/OCA 95Defining PreferencesTo enter link and location information:1. From the Main Menu, press Preferences.2

Pagina 5 - Certification Information

Setting Up the Instrument in Drift Mode96 FTB-5230S/-OCADefining Preferences3. Under System and link information, define the following parameters as n

Pagina 6

Setting Up the Instrument in Drift ModeOSA/OCA 97Defining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Pagina 7 - Analyzers

Setting Up the Instrument in Drift Mode98 FTB-5230S/-OCADefining Preferences2. Select the Comments tab.3. Enter your comments for the current trace. 4

Pagina 8 - Typical Applications

Setting Up the Instrument in Drift ModeOSA/OCA 99Defining PreferencesDefining Display ParametersThe application allows you to set display settings for

Pagina 9 - Post-Processing Application

Setting Up the Instrument in Drift Mode100 FTB-5230S/-OCADefining Preferences2. Select the Display tab.3. Select the spectral unit you want to work wi

Pagina 10 - IMPORTANT

Setting Up the Instrument in Drift ModeOSA/OCA 101Defining Preferences4. Select the label that will appear on the peaks in the graph, either the chann

Pagina 11 - 2 Safety Information

Setting Up the Instrument in Drift Mode102 FTB-5230S/-OCADefining PreferencesNote: The channel name and channel number cannot be shown at the same tim

Pagina 12 - 6 FTB-5230S/-OCA

Setting Up the Instrument in Drift ModeOSA/OCA 103Defining Preferences5. Select whether you want to show or hide the empty channels from the channel l

Pagina 13 - 3 Getting Started with Your

OSA/OCA 52 Safety InformationWARNINGDo not install or terminate fibers while a light source is active. Never look directly into a live fiber and ensur

Pagina 14 - 8 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode104 FTB-5230S/-OCADefining Preferences7. Select the background color scheme for the graph as desired. 8. Press

Pagina 15 - FTB-500 left panel

Setting Up the Instrument in Drift ModeOSA/OCA 105Defining PreferencesCustomizing Drift Results Table It is possible to select which results you would

Pagina 16 - 10 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode106 FTB-5230S/-OCADefining Preferences2. Select the Drift Results tab.3. Select which parameters you want to di

Pagina 17 - OSA/OCA 11

Setting Up the Instrument in Drift ModeOSA/OCA 107Setting Up Drift Analysis ParametersSetting Up Drift Analysis ParametersThis section presents the va

Pagina 18 - 12 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode108 FTB-5230S/-OCASetting Up Drift Analysis Parameters3. From any tab, press Import from Trace.4. Press OK to c

Pagina 19 - FTB-5230S

Setting Up the Instrument in Drift ModeOSA/OCA 109Setting Up Drift Analysis ParametersDefining General SettingsThe general analysis parameters for dri

Pagina 20 - Exiting the Application

Setting Up the Instrument in Drift Mode110 FTB-5230S/-OCASetting Up Drift Analysis Parameters2. Select the General tab.

Pagina 21

Setting Up the Instrument in Drift ModeOSA/OCA 111Setting Up Drift Analysis Parameters3. Under Default channel settings, define the following paramete

Pagina 22 - 16 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode112 FTB-5230S/-OCASetting Up Drift Analysis Parameters Channel width (GHz or nm): indicates the limit inside w

Pagina 23

Setting Up the Instrument in Drift ModeOSA/OCA 113Setting Up Drift Analysis ParametersPeak signal power: The peak signal power represents the maximum

Pagina 24 - Selecting a Test Mode

Safety Information6 FTB-5230S/-OCAYour instrument is a Class 1 laser product in compliance with standards IEC 60825-1: 2007 and 21 CFR 1040.10, except

Pagina 25

Setting Up the Instrument in Drift Mode114 FTB-5230S/-OCASetting Up Drift Analysis ParametersFifth order polynomial fit (Fit): The fifth order polyfit

Pagina 26 - Nulling Electrical Offsets

Setting Up the Instrument in Drift ModeOSA/OCA 115Setting Up Drift Analysis Parameters4. Under Global analysis parameters, define the following parame

Pagina 27

Setting Up the Instrument in Drift Mode116 FTB-5230S/-OCASetting Up Drift Analysis Parameters Wavelength offset (nm): indicates the offset value appl

Pagina 28

Setting Up the Instrument in Drift ModeOSA/OCA 117Setting Up Drift Analysis ParametersDefining Default Channel ThresholdsThe thresholds will be applie

Pagina 29 - Performing User Calibration

Setting Up the Instrument in Drift Mode118 FTB-5230S/-OCASetting Up Drift Analysis ParametersWhen thresholds are globally disabled, the results in the

Pagina 30 - 24 FTB-5230S/-OCA

Setting Up the Instrument in Drift ModeOSA/OCA 119Setting Up Drift Analysis ParametersYou can set your pass/fail threshold limits in different ways de

Pagina 31

Setting Up the Instrument in Drift Mode120 FTB-5230S/-OCASetting Up Drift Analysis ParametersTo define the default channel thresholds:1. From the Main

Pagina 32

Setting Up the Instrument in Drift ModeOSA/OCA 121Setting Up Drift Analysis Parameters2. Select the Thresholds tab.

Pagina 33 - Factors

Setting Up the Instrument in Drift Mode122 FTB-5230S/-OCASetting Up Drift Analysis Parameters3. Select the Activate all thresholds option to manually

Pagina 34 - 28 FTB-5230S/-OCA

Setting Up the Instrument in Drift ModeOSA/OCA 123Setting Up Drift Analysis ParametersManaging ChannelsTesting DWDM systems involves characterizing mu

Pagina 35

OSA/OCA 73 Getting Started with Your OSA/OCAInserting and Removing Test ModulesTo insert a module into the FTB-5230S/-OCA:1. Exit ToolBox and turn off

Pagina 36 - 30 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode124 FTB-5230S/-OCASetting Up Drift Analysis Parameters2. Select the Channels tab. 3. By default, the channel li

Pagina 37 - To save a user calibration:

Setting Up the Instrument in Drift ModeOSA/OCA 125Setting Up Drift Analysis Parameters4. Enter values in the boxes as explained below: Start range (n

Pagina 38 - 32 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode126 FTB-5230S/-OCASetting Up Drift Analysis Parameters Channel distance (nm or GHz): distance between the chan

Pagina 39

Setting Up the Instrument in Drift ModeOSA/OCA 127Setting Up Drift Analysis ParametersTo edit the parameters of a specific channel:1. From the Main Me

Pagina 40 - Using the Autonaming Feature

Setting Up the Instrument in Drift Mode128 FTB-5230S/-OCASetting Up Drift Analysis Parameters2. Select the Channels tab.

Pagina 41 - To customize the file name:

Setting Up the Instrument in Drift ModeOSA/OCA 129Setting Up Drift Analysis Parameters3. Select the channel or channels to be modified in the channel

Pagina 42

Setting Up the Instrument in Drift Mode130 FTB-5230S/-OCASetting Up Drift Analysis Parameters5. Modify the settings as needed. For more information ab

Pagina 43

Setting Up the Instrument in Drift ModeOSA/OCA 131Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing your test, you m

Pagina 44 - 38 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode132 FTB-5230S/-OCASetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main wi

Pagina 45

Setting Up the Instrument in Drift ModeOSA/OCA 133Setting Up Acquisition Parameters2. Select the acquisition type. 3. If you are performing an averagi

Pagina 46

Getting Started with Your OSA/OCA8 FTB-5230S/-OCAInserting and Removing Test Modules3. Take the module and place it so that the connector pins are at

Pagina 47 - WDM Mode

Setting Up the Instrument in Drift Mode134 FTB-5230S/-OCASetting Up Acquisition ParametersThe wavelength range covered within these bands of the spect

Pagina 48 - 42 FTB-5230S/-OCA

Setting Up the Instrument in Drift ModeOSA/OCA 135Setting Up Acquisition Parameters3. Select a sampling unit and enter a sampling count to configure a

Pagina 49

Setting Up the Instrument in Drift Mode136 FTB-5230S/-OCASetting Up Acquisition Parameters6. Select a location where the drift file should be saved. 7

Pagina 50 - 44 FTB-5230S/-OCA

Setting Up the Instrument in Drift ModeOSA/OCA 137Setting Up Acquisition Parameters8. To test your parameters, return to the Acquisition tab. Press Tr

Pagina 51

Setting Up the Instrument in Drift Mode138 FTB-5230S/-OCABuilding a Custom Drift MeasurementBuilding a Custom Drift MeasurementYou can build a drift m

Pagina 52

Setting Up the Instrument in Drift ModeOSA/OCA 139Building a Custom Drift MeasurementAcquisition start time Target WDM measurement acquisition start

Pagina 53

Setting Up the Instrument in Drift Mode140 FTB-5230S/-OCABuilding a Custom Drift MeasurementPower offset Target WDM measurement power offset differs f

Pagina 54

Setting Up the Instrument in Drift ModeOSA/OCA 141Building a Custom Drift MeasurementTo build a custom drift measurement:1. If you have not done so al

Pagina 55

Setting Up the Instrument in Drift Mode142 FTB-5230S/-OCABuilding a Custom Drift Measurement4. Select whether you want to import the analysis setup fr

Pagina 56

Setting Up the Instrument in Drift ModeOSA/OCA 143Building a Custom Drift Measurement5. Enter, or review if they were imported, the general details fo

Pagina 57

Getting Started with Your OSA/OCAOSA/OCA 9Inserting and Removing Test Modules7. While applying slight pressure to the module, turn the retaining screw

Pagina 58 - 52 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode144 FTB-5230S/-OCABuilding a Custom Drift Measurement7. If desired, adjust the threshold settings for your meas

Pagina 59

Setting Up the Instrument in Drift ModeOSA/OCA 145Building a Custom Drift Measurement8. Select which channels are to be included in the drift measurem

Pagina 60 - 54 FTB-5230S/-OCA

Setting Up the Instrument in Drift Mode146 FTB-5230S/-OCABuilding a Custom Drift Measurement9. Add one or more measurement files at this point, then c

Pagina 61

Setting Up the Instrument in Drift ModeOSA/OCA 147Building a Custom Drift Measurement10. Before starting the measurement process, you can select what

Pagina 63 - Customizing WDM Results Table

OSA/OCA 1497 Setting Up Your OCAPerforming a spectral analysis with your Optical Channel Analyzer (OCA) is simple and can be achieved after selecting

Pagina 64

Setting Up Your OCA150 FTB-5230S/-OCATo set parameters in the Acquisition tab:1. From the main window, select the Acquisition tab.2. Select the acquis

Pagina 65 - OSA/OCA 59

Setting Up Your OCAOSA/OCA 151To create a channel plan and use it as one of the Custom channel plans:1. Start the post-processing application on a com

Pagina 66 - 60 FTB-5230S/-OCA

Setting Up Your OCA152 FTB-5230S/-OCA7. In Windows Explorer, rename the extension of your file from .osawdmsetup to .ocawdmsetup.8. Using a USB key or

Pagina 67 - OSA/OCA 61

OSA/OCA 1538 Starting a MeasurementBefore starting a measurement you must select and configure a test mode. You will find the instructions to select a

Pagina 68

Getting Started with Your OSA/OCA10 FTB-5230S/-OCAInserting and Removing Test ModulesTo remove a module from the FTB-500:1. Exit ToolBox and turn off

Pagina 70 - 64 FTB-5230S/-OCA

OSA/OCA 1559 Managing Files and Test ConfigurationsUsing the Discover FeatureThe Discover feature allows you to start a measurement procedure to autom

Pagina 71 - OSA/OCA 65

Managing Files and Test Configurations156 FTB-5230S/-OCAUsing the Discover FeatureDiscover analysis parameters are established as follows: The acquis

Pagina 72 - 66 FTB-5230S/-OCA

Managing Files and Test ConfigurationsOSA/OCA 157Using the Discover FeatureTo start an automatic setup measurement:Note: You cannot do a setup measure

Pagina 73

Managing Files and Test Configurations158 FTB-5230S/-OCAManaging Measurement FilesManaging Measurement FilesThe application allows you to manage the m

Pagina 74 - 68 FTB-5230S/-OCA

Managing Files and Test ConfigurationsOSA/OCA 159Managing Measurement FilesTo o pe n a f il e:1. From the Main Menu, Press File, and then press Open (

Pagina 75 - OSA/OCA 69

Managing Files and Test Configurations160 FTB-5230S/-OCAManaging Measurement Files3. Scroll through the list and select a trace to open.4. If you are

Pagina 76 - Defining Global Thresholds

Managing Files and Test ConfigurationsOSA/OCA 161Managing Measurement FilesTo c le ar a fi le :1. From the Main Menu, Press File. (for the OCA applic

Pagina 77

Managing Files and Test Configurations162 FTB-5230S/-OCAOpening Other Test Mode Files in WDM ModeOpening Other Test Mode Files in WDM ModeYour applica

Pagina 78 - To define global thresholds:

Managing Files and Test ConfigurationsOSA/OCA 163Managing FavoritesManaging FavoritesFavorites are configuration files that contain all of the paramet

Pagina 79 - OSA/OCA 73

Getting Started with Your OSA/OCAOSA/OCA 11Inserting and Removing Test Modules5. Hold the module by its sides or by the handle (NOT by the connector)

Pagina 80 - Defining Default Thresholds

Managing Files and Test Configurations164 FTB-5230S/-OCAManaging Favorites2. Select the Favorites tab.

Pagina 81

Managing Files and Test ConfigurationsOSA/OCA 165Managing Favorites3. To apply the settings from a favorite file to the current analysis setup, select

Pagina 82 - To define default Thresholds:

Managing Files and Test Configurations166 FTB-5230S/-OCAManaging FavoritesTo save a test configuration:1. From the Main Menu, press Analysis Setup. OR

Pagina 83 - OSA/OCA 77

Managing Files and Test ConfigurationsOSA/OCA 167Managing Favorites2. Select the Favorites tab.3. To save an analysis setup to a file, press Save As.

Pagina 84

Managing Files and Test Configurations168 FTB-5230S/-OCAManaging Favorites4. In the Save As window, enter a file name and press Save. The file will be

Pagina 85

Managing Files and Test ConfigurationsOSA/OCA 169Managing Favorites2. Select the Favorites tab.3. Press Import to import an analysis setup from a file

Pagina 86

Managing Files and Test Configurations170 FTB-5230S/-OCAManaging Favorites4. From the Import window, select the file you want to import and press Open

Pagina 87 - OSA/OCA 81

Managing Files and Test ConfigurationsOSA/OCA 171Managing FavoritesTo delete a test configuration:1. From the Main Menu, press Analysis Setup. ORFrom

Pagina 88 - 82 FTB-5230S/-OCA

Managing Files and Test Configurations172 FTB-5230S/-OCAManaging Favorites2. Select the Favorites tab.

Pagina 89 - OSA/OCA 83

Managing Files and Test ConfigurationsOSA/OCA 173Importing a Configuration from the Current Trace3. To delete a configuration file from the favorites

Pagina 90 - 4. Press Edit Selection

Getting Started with Your OSA/OCA12 FTB-5230S/-OCAStarting the OSA/OCA ApplicationStarting the OSA/OCA ApplicationYour FTB-5230S/-OCA Optical Spectrum

Pagina 92 - 86 FTB-5230S/-OCA

OSA/OCA 17510 Managing ResultsEach test mode has its own results tabs, where you can view the trace details, channel results and global results for al

Pagina 93 - OSA/OCA 87

Managing Results176 FTB-5230S/-OCAManaging WDM Test ResultsManaging WDM Test ResultsThe application allows you to view and manage your WDM test result

Pagina 94 - 88 FTB-5230S/-OCA

Managing ResultsOSA/OCA 177Managing WDM Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 153 for details on how to perfor

Pagina 95 - OSA/OCA 89

Managing Results178 FTB-5230S/-OCAManaging WDM Test ResultsThe selected peak indicator, a small red inverted triangle ( ) points down at the top of th

Pagina 96

Managing ResultsOSA/OCA 179Managing WDM Test ResultsThe noise level for a channel is indicated by a dotted line under the selected peak. The width of

Pagina 97 - Drift Mode

Managing Results180 FTB-5230S/-OCAManaging WDM Test ResultsYou can move the cursor to modify the peak detection level for the current measurement.Each

Pagina 98 - 92 FTB-5230S/-OCA

Managing ResultsOSA/OCA 181Managing WDM Test ResultsResults TabIn the Results tab, each channel will be represented for both the active and reference

Pagina 99

Managing Results182 FTB-5230S/-OCAManaging WDM Test ResultsChannel Results TabThe application allows you to view the complete information about the pa

Pagina 100 - Defining Preferences

Managing ResultsOSA/OCA 183Managing WDM Test Results2. Select a row from the Results tab to view the results for this channel.Note: Values displayed i

Pagina 101

Getting Started with Your OSA/OCAOSA/OCA 13Starting the OSA/OCA Application2. Click the corresponding button in the Module Applications box.The main w

Pagina 102

Managing Results184 FTB-5230S/-OCAManaging WDM Test ResultsGlobal Results TabThe application allows you to view the global results of the current meas

Pagina 103 - To enter comments:

Managing ResultsOSA/OCA 185Managing Drift Test ResultsManaging Drift Test ResultsThe application allows you to view and manage your drift test results

Pagina 104 - 98 FTB-5230S/-OCA

Managing Results186 FTB-5230S/-OCAManaging Drift Test ResultsYou can select a channel directly from the dashboard or from the Channel History tab. For

Pagina 105 - Defining Display Parameters

Managing ResultsOSA/OCA 187Managing Drift Test ResultsThe dashboard shows a global status (all channels) for each parameter. This global status is set

Pagina 106 - 2. Select the Display tab

Managing Results188 FTB-5230S/-OCAManaging Drift Test ResultsThe dashboard displays a channel status (all parameters) for a given channel. This channe

Pagina 107

Managing ResultsOSA/OCA 189Managing Drift Test ResultsChannel Graph TabThe Channel Graph tab displays three different graphs for the selected channel.

Pagina 108 - 102 FTB-5230S/-OCA

Managing Results190 FTB-5230S/-OCAManaging Drift Test ResultsChannel History TabThe channel history table shows channel results for the active trace.

Pagina 109

Managing ResultsOSA/OCA 191Managing Drift Test ResultsFor each of the above parameters, the following results are displayed: Reference: channel refer

Pagina 110 - 104 FTB-5230S/-OCA

Managing Results192 FTB-5230S/-OCAManaging Drift Test ResultsWDM Graph TabThe WDM Graph tab allows you to view the spectrum of the active trace for th

Pagina 111

Managing ResultsOSA/OCA 193Managing Drift Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 153 for details on how to perf

Pagina 112 - 106 FTB-5230S/-OCA

ii FTB-5230S/-OCACopyright © 2013 EXFO Inc. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transm

Pagina 113 - OSA/OCA 107

Getting Started with Your OSA/OCA14 FTB-5230S/-OCAExiting the ApplicationNote: The illustrations in this user guide may differ slightly from those on

Pagina 114 - 108 FTB-5230S/-OCA

Managing Results194 FTB-5230S/-OCAManaging Drift Test ResultsChannel Results TabWhen you select a channel in the Channel History tab, the Channel Resu

Pagina 115

Managing ResultsOSA/OCA 195Managing Drift Test Results2. Select a channel from the Channel History tab to view the channel results for the selected ch

Pagina 116 - 2. Select the General tab

Managing Results196 FTB-5230S/-OCAManaging OCA Test ResultsManaging OCA Test ResultsThe application allows you to view and manage your OCA test result

Pagina 117 - OSA/OCA 111

Managing ResultsOSA/OCA 197Managing OCA Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 153 for details on how to perfor

Pagina 118 - 112 FTB-5230S/-OCA

Managing Results198 FTB-5230S/-OCAManaging OCA Test ResultsThe graph will display peak indicators for all the channels found by the application with a

Pagina 119 - OSA/OCA 113

Managing ResultsOSA/OCA 199Managing OCA Test ResultsThe graph will also display a dotted line across the full spectral width corresponding to the peak

Pagina 120 - 114 FTB-5230S/-OCA

Managing Results200 FTB-5230S/-OCAManaging OCA Test ResultsResults TabThe results table shows the current channel results for the trace and the total

Pagina 121 - OSA/OCA 115

Managing ResultsOSA/OCA 201Adjusting the Display SizeAdjusting the Display SizeYour application allows you to toggle the view of your main window. You

Pagina 122 - 116 FTB-5230S/-OCA

Managing Results202 FTB-5230S/-OCAAdjusting the Display SizeFor 100 % lower tabs view, press .

Pagina 123 - OSA/OCA 117

Managing ResultsOSA/OCA 203Viewing WDM Graph in Full-Screen ModeViewing WDM Graph in Full-Screen ModeThe full-screen mode allows you to see the WDM gr

Pagina 124 - 118 FTB-5230S/-OCA

OSA/OCA 154 Preparing Your OSA/OCA for a TestCleaning and Connecting Optical FibersTo connect the fiber-optic cable to the port:1. Inspect the fiber u

Pagina 125 - Threshold Limit Definition

Managing Results204 FTB-5230S/-OCAUsing Zoom ControlsUsing Zoom ControlsUse the zoom controls to change the scale of the trace display.You can zoom in

Pagina 126 - 120 FTB-5230S/-OCA

Managing ResultsOSA/OCA 205Using Zoom ControlsTo view specific portions of the graph: You can define which portion of the graph will be visible by pr

Pagina 127 - OSA/OCA 121

Managing Results206 FTB-5230S/-OCAManaging MarkersManaging MarkersYou can use markers to perform manual measurements and verification directly on the

Pagina 128 - 122 FTB-5230S/-OCA

Managing ResultsOSA/OCA 207Managing MarkersYou can also use the automated marker positioning to place the markers around a specific channel peak. The

Pagina 129 - Managing Channels

Managing Results208 FTB-5230S/-OCAManaging MarkersTo manually enter a marker position value:1. If you have not done so already, press the button on

Pagina 130 - 2. Select the Channels tab

Managing ResultsOSA/OCA 209Managing MarkersAs markers C and D appear on the graph, the power difference between the markers (C-D) related to the horiz

Pagina 131 - OSA/OCA 125

Managing Results210 FTB-5230S/-OCAManaging Trace InformationManaging Trace InformationAfter acquiring a trace, you might want to view details about th

Pagina 132

Managing ResultsOSA/OCA 211Managing Trace InformationTo edit general information:1. From the main window, select the Trace Info. tab.2. Press Trace Id

Pagina 133 - OSA/OCA 127

Managing Results212 FTB-5230S/-OCAManaging Trace InformationTo edit trace information:Note: This feature is not available in the OCA application.1. Fr

Pagina 134

Managing ResultsOSA/OCA 213Managing Trace InformationTo edit comments:Note: This feature is not available in the OCA application.1. From the main wind

Pagina 135 - OSA/OCA 129

Preparing Your OSA/OCA for a Test16 FTB-5230S/-OCACleaning and Connecting Optical Fibers3. Carefully align the connector and port to prevent the fiber

Pagina 136 - 130 FTB-5230S/-OCA

Managing Results214 FTB-5230S/-OCAGenerating ReportsGenerating ReportsAfter performing any acquisition, you can generate a report for the current acqu

Pagina 137 - OSA/OCA 131

Managing ResultsOSA/OCA 215Generating Reports4. From the Save as type list, select the format for your report.5. Press Save. The report will be added

Pagina 139 - OSA/OCA 133

OSA/OCA 21711 MaintenanceTo help ensure long, trouble-free operation: Always inspect fiber-optic connectors before using them and clean them if neces

Pagina 140

Maintenance218 FTB-5230S/-OCACleaning EUI ConnectorsTo clean EUI connectors:1. Remove the EUI from the instrument to expose the connector baseplate an

Pagina 141 - OSA/OCA 135

MaintenanceOSA/OCA 219Cleaning EUI Connectors6. Clean the ferrule in the connector port as follows:6a. Deposit one drop of isopropyl alcohol on a lint

Pagina 142 - 136 FTB-5230S/-OCA

Maintenance220 FTB-5230S/-OCARecalibrating the UnitRecalibrating the UnitEXFO manufacturing and service center calibrations are based on the ISO/IEC 1

Pagina 143 - OSA/OCA 137

MaintenanceOSA/OCA 221Recycling and Disposal (Applies to European Union Only)To help you with calibration follow-up, EXFO provides a special calibrati

Pagina 145 - OSA/OCA 139

OSA/OCA 22312 TroubleshootingViewing Online DocumentationIn addition to the online help available from the application, you will also find a printable

Pagina 146 - Compatibility

Preparing Your OSA/OCA for a TestOSA/OCA 17Installing the EXFO Universal Interface (EUI)Installing the EXFO Universal Interface (EUI)The EUI fixed bas

Pagina 147 - OSA/OCA 141

Troubleshooting224 FTB-5230S/-OCAContacting the Technical Support GroupContacting the Technical Support GroupTo obtain after-sales service or technica

Pagina 148 - 142 FTB-5230S/-OCA

TroubleshootingOSA/OCA 225Contacting the Technical Support GroupTo view the information about the product:From the Main Menu, press .

Pagina 149 - 6. Click Next

Troubleshooting226 FTB-5230S/-OCATransportationTransportationMaintain a temperature range within specifications when transporting the unit. Transporta

Pagina 150 - 144 FTB-5230S/-OCA

TroubleshootingOSA/OCA 227TransportationThe following image show the GP-10-055 case with the module inside.Module

Pagina 152 - 146 FTB-5230S/-OCA

OSA/OCA 22913 WarrantyGeneral InformationEXFO Inc. (EXFO) warrants this equipment against defects in material and workmanship for a period of 1 year f

Pagina 153 - OSA/OCA 147

Warranty230 FTB-5230S/-OCALiabilityLiabilityEXFO shall not be liable for damages resulting from the use of the product, nor shall be responsible for a

Pagina 154

WarrantyOSA/OCA 231ExclusionsExclusionsEXFO reserves the right to make changes in the design or construction of any of its products at any time withou

Pagina 155 - 7 Setting Up Your OCA

Warranty232 FTB-5230S/-OCAService and RepairsService and RepairsEXFO commits to providing product service and repair for five years following the date

Pagina 156 - Setting Up Your OCA

WarrantyOSA/OCA 233EXFO Service Centers WorldwideEXFO Service Centers WorldwideIf your product requires servicing, contact your nearest authorized ser

Pagina 157 - OSA/OCA 151

Preparing Your OSA/OCA for a Test18 FTB-5230S/-OCASelecting a Test ModeSelecting a Test ModeYour module gives you different ways to test all your DWDM

Pagina 159 - 8 Starting a Measurement

OSA/OCA 235A Technical SpecificationsIMPORTANTThe following technical specifications can change without notice. The information presented in this sect

Pagina 161 - Configurations

IndexOSA/OCA 237IndexAacquisitionaveraging... 86, 131real-time ... 86, 149sin

Pagina 162

Index238 FTB-5230S/-OCADdefaultchannel, activating ... 64, 111thresholds... 74deleting

Pagina 163 - Using the Discover Feature

IndexOSA/OCA 239inserting a module ... 7integrated power ... 65, 112ITU grid ...

Pagina 164 - Managing Measurement Files

Index240 FTB-5230S/-OCAnoise... 66, 113resolution bandwidth ... 68Pparametersdispl

Pagina 165

IndexOSA/OCA 241flatness ... 73peak... 113signal to n

Pagina 166 - 160 FTB-5230S/-OCA

Index242 FTB-5230S/-OCAWDManalysis parameters ... 59general settings ... 62opening files

Pagina 167

NOTICE抩⛙CHINESE REGULATION ON RESTRICTION OF HAZARDOUS SUBSTANCES₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩NAMES AND CONTENTS OF THE TOXIC OR HAZARDOUS SUBSTANCES OR ELEMENTS CONT

Pagina 168 - 162 FTB-5230S/-OCA

Preparing Your OSA/OCA for a TestOSA/OCA 19Selecting a Test Mode2. Select the desired test mode.Once you select the mode, you will notice a against

Pagina 169 - Managing Favorites

MARKING REQUIREMENTS㪖㽷尐㻑Productℶ❐Environmental protection use period (years)䘾⬒≬㔳∎䞷㦮棟 ( ㄃ )Logo㪖㉦This Exfo product㦻 EXFO ℶ❐10Batterya䟄㻯aa. If applicabl

Pagina 170 - 2. Select the Favorites tab

www.EXFO.com · [email protected]/N: 1065480CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADATel.: 1 418 683-0211 · Fax: 1 418 683-

Pagina 171

Preparing Your OSA/OCA for a Test20 FTB-5230S/-OCANulling Electrical OffsetsNulling Electrical OffsetsThe offset nulling process provides a zero-power

Pagina 172 - To save a test configuration:

Preparing Your OSA/OCA for a TestOSA/OCA 21Nulling Electrical Offsets2. Disconnect any incoming signal to obtain an optimal accuracy.3. Press Nulling.

Pagina 173

Preparing Your OSA/OCA for a Test22 FTB-5230S/-OCAPerforming User CalibrationPerforming User CalibrationCalibrating your module can help you achieve b

Pagina 174 - 168 FTB-5230S/-OCA

Preparing Your OSA/OCA for a TestOSA/OCA 23Performing User CalibrationNote: The user calibration feature is not available in the offline version of th

Pagina 175

OSA/OCA iiiContentsCertification Information ...

Pagina 176 - 170 FTB-5230S/-OCA

Preparing Your OSA/OCA for a Test24 FTB-5230S/-OCAPerforming User Calibration3. Select the Calibration tab. Note: You cannot edit the power or wavelen

Pagina 177

Preparing Your OSA/OCA for a TestOSA/OCA 25Performing User Calibration4. If user correction factors are in the system, press Clear User Correction Fac

Pagina 178

Preparing Your OSA/OCA for a Test26 FTB-5230S/-OCAPerforming User Calibration6. Note the measurements to a .txt file using the following format: The

Pagina 179 - Using a Restore Point

Preparing Your OSA/OCA for a TestOSA/OCA 27Performing User Calibration8. Back in the Calibration tab on your unit, load the file using Load Factors.9.

Pagina 180

Preparing Your OSA/OCA for a Test28 FTB-5230S/-OCAPerforming User CalibrationThe calibration values will replace the Correction factors list in the An

Pagina 181 - 10 Managing Results

Preparing Your OSA/OCA for a TestOSA/OCA 29Performing User Calibration10. Press Write to Module to apply the modified calibration values to the module

Pagina 182 - Managing WDM Test Results

Preparing Your OSA/OCA for a Test30 FTB-5230S/-OCAPerforming User Calibration11. To verify that the calibration changes are properly applied to the mo

Pagina 183

Preparing Your OSA/OCA for a TestOSA/OCA 31Performing User CalibrationTo save a user calibration:1. From the Main Menu, press Analysis Setup.

Pagina 184 - 178 FTB-5230S/-OCA

Preparing Your OSA/OCA for a Test32 FTB-5230S/-OCAPerforming User Calibration2. Select the Calibration tab.

Pagina 185

Preparing Your OSA/OCA for a TestOSA/OCA 33Performing User Calibration3. Press Save Factors, to save the modified user calibration values.

Pagina 186 - 180 FTB-5230S/-OCA

iv FTB-5230S/-OCA9 Managing Files and Test Configurations ...155Using the Discover Feature ...

Pagina 187 - Results Tab

Preparing Your OSA/OCA for a Test34 FTB-5230S/-OCAUsing the Autonaming FeatureUsing the Autonaming FeatureDefining a file autonaming format will allow

Pagina 188 - Channel Results Tab

Preparing Your OSA/OCA for a TestOSA/OCA 35Using the Autonaming FeatureTo customize the file name:1. From the Main Menu, press Preferences.

Pagina 189

Preparing Your OSA/OCA for a Test36 FTB-5230S/-OCAUsing the Autonaming Feature2. Select the File Name tab.3. Select which parameters you want to inclu

Pagina 190 - Global Results Tab

Preparing Your OSA/OCA for a TestOSA/OCA 37Using the Autonaming Feature4. Press the up or down arrows to change the order in which the field values wi

Pagina 191 - Managing Drift Test Results

Preparing Your OSA/OCA for a Test38 FTB-5230S/-OCAUsing the Autonaming FeatureTo customize the file name in the OCA application:1. From the main windo

Pagina 192 -  Signal power

Preparing Your OSA/OCA for a TestOSA/OCA 39Using the Autonaming Feature4. Select the Use as template option.5. Press OK to confirm your choice.The new

Pagina 194

OSA/OCA 415 Setting Up the Instrument in WDM ModeNote: This mode is not available in the OCA application.Before performing a spectral analysis in the

Pagina 195 - Channel Graph Tab

Setting Up the Instrument in WDM Mode42 FTB-5230S/-OCAYou can set up your unit in different manners, depending on your testing needs. The preferred w

Pagina 196 - Channel History Tab

Setting Up the Instrument in WDM ModeOSA/OCA 43Defining PreferencesDefining PreferencesThe preferences window allows you to set general information an

Pagina 197 - OSA/OCA 191

Certification InformationOSA/OCA vCertification InformationNorth America Regulatory StatementThis unit was certified by an agency approved in both Can

Pagina 198 - WDM Graph Tab

Setting Up the Instrument in WDM Mode44 FTB-5230S/-OCADefining Preferences2. Select the General tab.3. Define the general parameters as needed.4. Pres

Pagina 199

Setting Up the Instrument in WDM ModeOSA/OCA 45Defining PreferencesTo enter link and location information:1. From the Main Menu, press Preferences.2.

Pagina 200

Setting Up the Instrument in WDM Mode46 FTB-5230S/-OCADefining Preferences3. Under System and link information, define the following parameters as nee

Pagina 201

Setting Up the Instrument in WDM ModeOSA/OCA 47Defining Preferences4. Under Location Information, define the following parameters as needed: Network

Pagina 202 - Managing OCA Test Results

Setting Up the Instrument in WDM Mode48 FTB-5230S/-OCADefining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Pagina 203 -  X axis: wavelength in nm

Setting Up the Instrument in WDM ModeOSA/OCA 49Defining Preferences2. Select the Comments tab.3. Enter your comments for the current trace. 4. Press O

Pagina 204 - 198 FTB-5230S/-OCA

Setting Up the Instrument in WDM Mode50 FTB-5230S/-OCADefining PreferencesDefining Display ParametersThe application allows you to set display setting

Pagina 205

Setting Up the Instrument in WDM ModeOSA/OCA 51Defining Preferences2. Select the Display tab.3. Select the spectral unit you want to work with, either

Pagina 206

Setting Up the Instrument in WDM Mode52 FTB-5230S/-OCADefining Preferences4. Select the label that will appear on the peaks in the graph, either the c

Pagina 207 - Adjusting the Display Size

Setting Up the Instrument in WDM ModeOSA/OCA 53Defining PreferencesNote: The channel name and channel number cannot be shown at the same time.Channel

Pagina 209

Setting Up the Instrument in WDM Mode54 FTB-5230S/-OCADefining Preferences5. Select whether you want to show or hide the empty channels from the chann

Pagina 210 - Using Zoom Controls

Setting Up the Instrument in WDM ModeOSA/OCA 55Defining Preferences6. Select whether you want to show the horizontal markers or the integrated power a

Pagina 211

Setting Up the Instrument in WDM Mode56 FTB-5230S/-OCADefining Preferences7. Select the background color scheme for the graph as desired. 8. Press OK

Pagina 212 - Managing Markers

Setting Up the Instrument in WDM ModeOSA/OCA 57Defining PreferencesCustomizing WDM Results TableIt is possible to select which results you would like

Pagina 213 - OSA/OCA 207

Setting Up the Instrument in WDM Mode58 FTB-5230S/-OCADefining Preferences2. Select the WDM Results tab.3. Select which parameters you want to display

Pagina 214

Setting Up the Instrument in WDM ModeOSA/OCA 59Setting Up WDM Analysis Parameters /f : deviation of the spectral center of mass for the peak in tha

Pagina 215

Setting Up the Instrument in WDM Mode60 FTB-5230S/-OCASetting Up WDM Analysis ParametersTo import the parameters from the current trace:1. Make sure t

Pagina 216 - Managing Trace Information

Setting Up the Instrument in WDM ModeOSA/OCA 61Setting Up WDM Analysis Parameters3. From any tab, press Import from Trace.4. Press OK to confirm the c

Pagina 217

Setting Up the Instrument in WDM Mode62 FTB-5230S/-OCASetting Up WDM Analysis ParametersDefining General SettingsThe general analysis parameters for W

Pagina 218 - 212 FTB-5230S/-OCA

Setting Up the Instrument in WDM ModeOSA/OCA 63Setting Up WDM Analysis Parameters2. Select the General tab.

Pagina 219

OSA/OCA 11 Introducing the FTB-5230S/-OCA Optical Spectrum and Channel AnalyzersThe FTB-5230S Optical Spectrum Analyzer (OSA) is designed to measure

Pagina 220 - Generating Reports

Setting Up the Instrument in WDM Mode64 FTB-5230S/-OCASetting Up WDM Analysis Parameters3. Under Default channel settings, define the following parame

Pagina 221

Setting Up the Instrument in WDM ModeOSA/OCA 65Setting Up WDM Analysis Parameters Channel width (GHz or nm): indicates the limit inside which the pow

Pagina 222

Setting Up the Instrument in WDM Mode66 FTB-5230S/-OCASetting Up WDM Analysis ParametersPeak signal power: The peak signal power represents the maximu

Pagina 223 - 11 Maintenance

Setting Up the Instrument in WDM ModeOSA/OCA 67Setting Up WDM Analysis Parameters Fifth order polynomial fit (Fit): The fifth order polyfit method ca

Pagina 224 - Cleaning EUI Connectors

Setting Up the Instrument in WDM Mode68 FTB-5230S/-OCASetting Up WDM Analysis Parameters4. Under Global analysis parameters, define the following para

Pagina 225

Setting Up the Instrument in WDM ModeOSA/OCA 69Setting Up WDM Analysis Parameters Wavelength offset (nm): indicates the offset value applied on the w

Pagina 226 - Recalibrating the Unit

Setting Up the Instrument in WDM Mode70 FTB-5230S/-OCASetting Up WDM Analysis ParametersDefining Global ThresholdsAny change you make to the global th

Pagina 227 - Recycling and Disposal

Setting Up the Instrument in WDM ModeOSA/OCA 71Setting Up WDM Analysis ParametersYou can set your pass/fail threshold limits in different ways dependi

Pagina 228

Setting Up the Instrument in WDM Mode72 FTB-5230S/-OCASetting Up WDM Analysis ParametersTo define global thresholds:1. From the Main Menu, press Analy

Pagina 229 - 12 Troubleshooting

Setting Up the Instrument in WDM ModeOSA/OCA 73Setting Up WDM Analysis Parameters3. Select the Activate all thresholds option to manually set the glob

Pagina 230 - 224 FTB-5230S/-OCA

Introducing the FTB-5230S/-OCA Optical Spectrum and Channel Analyzers2 FTB-5230S/-OCAModelsModelsThe OSA/OCA comes in different models: The FTB-5230S

Pagina 231 - From the Main Menu, press

Setting Up the Instrument in WDM Mode74 FTB-5230S/-OCASetting Up WDM Analysis Parameters5. Press OK to save the changes and close the window, or press

Pagina 232

Setting Up the Instrument in WDM ModeOSA/OCA 75Setting Up WDM Analysis ParametersYou can set your pass/fail threshold limits in different ways dependi

Pagina 233

Setting Up the Instrument in WDM Mode76 FTB-5230S/-OCASetting Up WDM Analysis ParametersTo define default Thresholds:1. From the Main Menu, press Anal

Pagina 234

Setting Up the Instrument in WDM ModeOSA/OCA 77Setting Up WDM Analysis Parameters2. Select the Default Thresholds tab. 3. Enter values in the boxes as

Pagina 235 - 13 Warranty

Setting Up the Instrument in WDM Mode78 FTB-5230S/-OCASetting Up WDM Analysis ParametersManaging ChannelsTesting DWDM systems involves characterizing

Pagina 236 - Liability

Setting Up the Instrument in WDM ModeOSA/OCA 79Setting Up WDM Analysis Parameters2. Select the Channels tab. 3. By default, the channel list is empty.

Pagina 237

Setting Up the Instrument in WDM Mode80 FTB-5230S/-OCASetting Up WDM Analysis Parameters4. Enter values in the boxes as explained below: Start range

Pagina 238 - Service and Repairs

Setting Up the Instrument in WDM ModeOSA/OCA 81Setting Up WDM Analysis Parameters Starting value: sets the increment starting value for the channel n

Pagina 239 - OSA/OCA 233

Setting Up the Instrument in WDM Mode82 FTB-5230S/-OCASetting Up WDM Analysis ParametersTo edit the parameters of a specific channel:1. From the Main

Pagina 240

Setting Up the Instrument in WDM ModeOSA/OCA 83Setting Up WDM Analysis Parameters2. Select the Channels tab. 3. Select the channel or channels to be m

Pagina 241 - A Technical Specifications

Introducing the FTB-5230S/-OCA Optical Spectrum and Channel AnalyzersOSA/OCA 3Optional Software PackagesOptional Software PackagesOptional software op

Pagina 242

Setting Up the Instrument in WDM Mode84 FTB-5230S/-OCASetting Up WDM Analysis Parameters4. Press Edit Selection.

Pagina 243 - OSA/OCA 237

Setting Up the Instrument in WDM ModeOSA/OCA 85Setting Up WDM Analysis Parameters5. Modify the settings as needed. For more information about the sett

Pagina 244

Setting Up the Instrument in WDM Mode86 FTB-5230S/-OCASetting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing your test, y

Pagina 245 - OSA/OCA 239

Setting Up the Instrument in WDM ModeOSA/OCA 87Setting Up Acquisition ParametersBefore performing measurements on an optical spectrum, you must select

Pagina 246

Setting Up the Instrument in WDM Mode88 FTB-5230S/-OCASetting Up Acquisition Parameters3. If you are performing an averaging type acquisition, enter t

Pagina 247 - OSA/OCA 241

Setting Up the Instrument in WDM ModeOSA/OCA 89Setting Up Acquisition ParametersThe wavelength range covered within these bands of the spectra are lis

Pagina 249 - ₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩

OSA/OCA 916 Setting Up the Instrument in Drift ModeNote: This mode is not available in the OCA application.Before performing a spectral analysis in th

Pagina 250 - MARKING REQUIREMENTS

Setting Up the Instrument in Drift Mode92 FTB-5230S/-OCAYou can set up your unit in different manners, depending on your testing needs. The preferred

Pagina 251

Setting Up the Instrument in Drift ModeOSA/OCA 93Defining PreferencesDefining PreferencesThe preferences window allows you to set general information

Comentarios a estos manuales

Sin comentarios