Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 199

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 251
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 198
Managing Results
OSA/OCA 193
Managing Drift Test Results
When the acquisition is taken (see Starting a Measurement on page 153 for
details on how to perform a test), the active trace will be displayed in the
tab with information along the following axis values:
X axis: wavelength in nm or frequency in THz.
Y axis: optical power expressed in dBm, as measured in the optical
resolution bandwidth (RBW) of the OSA/OCA. This reference RBW is
shown at the bottom of the graph.
The graph will display peak indicators for all the channels found by the
application with a red vertical line over the peaks to indicate the peak
position.
A blue horizontal bar ( ) will be displayed on the top of a channel if it
does not overlap with another channel. If the channel overlaps with
another channel, the horizontal bar will be yellow ( ).
Vista de pagina 198
1 2 ... 194 195 196 197 198 199 200 201 202 203 204 ... 250 251

Comentarios a estos manuales

Sin comentarios