Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual de usuario Pagina 138

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 251
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 137
Setting Up the Instrument in Drift Mode
132 FTB-5230S/-OCA
Setting Up Acquisition Parameters
Setting Up Acquisition Parameters
Before performing your test, you must set the acquisition type and
parameters from the Acquisition tab and other parameters from the Drift
Settings tab.
There are three types of acquisitions in Drift mode: single, averaging and
InBand.
Single: Spectral measurement is performed once. The results are
displayed according to this measurement.
Averaging: Spectral measurements are performed based on the
number of scans that you have entered for this parameter. The trace
will be displayed after each acquisition and averaged with the previous
traces.
Before performing measurements on an optical spectrum, you must select
the wavelength/frequency range to use. You can perform the scan on the
full range, on spectral bands, or select a custom range.
Note: The shorter the wavelength or frequency range, the faster the acquisition.
You can configure the delay, sampling and total duration for a drift
measurement. You can also configure the drift files name and select a
location where it should be saved.
The application allows you to perform a trial scan while setting up the drift
measurement.
Vista de pagina 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 250 251

Comentarios a estos manuales

Sin comentarios