Troubleshooting
78 FTB-3930
Solving Common Problems
Unable to establish FASTEST
communication.
Remote unit’s FASTEST
port is in use.
Remote unit is not
compatible.
Selected port on master
unit differs from actual
port to which fiber is
connected.
Compatibility set to
FOT-930 / FTB-3930
but remote unit is not
an FOT-930 or
FTB-3930.
Wait until FASTEST
completes, turn source
and ORL meter off or exit
Probe mode (FOT-930).
Make sure remote unit
isan FOT-930, FTB-3930,
FOT-920 or FTB-3920.
Set the port (SM or MM)
correctlyin Setup (F
ASTEST
tab) on master unit.
Set the compatibility to
FOT-920 / FTB-3920 in
Setup (F
ASTEST tab) on
master unit.
During F
ASTEST, you get a
message saying that
loopback reference was
not performed.
Loopback reference was
selected, but not performed
(or reference is outdated).
Perform a loopback reference
before F
ASTEST measurement.
ORL values are inaccurate
for short fibers at 1310 nm
during a F
ASTEST.
— Measure ORL manually.
Unable to communicate
with an FOT-920 using the
talk set.
Talk sets of the FOT-920 and
FOT-930/FTB-3930 are not
compatible.
Use text messaging instead.
Unable to establish
connection with
compatible talk set.
Probe mode is activated on
remote unit(FOT-930 only).
Deactivate Probe mode.
Problem Possible Cause Solution