Exfo FTB-3930 MultiTest Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 101
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
Performing Automated IL/ORL/Length Measurements (FasTesT)
36 FTB-3930
Setting Up the FasTest
Setting Up the FasTest
You can configure the FASTEST on the master unit only. The remote unit will
automatically adapt to these parameters (F
ASTEST parameters on remote
unit are ignored).
Setup includes the following elements:
Port and DUT (fiber) type: multimode FASTEST will not include ORL.
Compatibility: select FOT-930 / FTB-3930 for fast, two- or
three-wavelength testing including ORL (it requires two
FOT-930/FTB-3930). Use FOT-920 / FTB-3920 when other unit is an
FOT-920 or FTB-3920.
Mode/wavelengths: select one or more wavelengths for the FASTEST.
Depending on your choice, the F
ASTEST will include loss and/or ORL
measurements. Selecting FTTx Custom or FTTx All allows you to
define upstream and downstream wavelengths.
Unit location: in FTTx mode, you specify if the master unit is closer to
the CO (or to the premises) than the remote unit.
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 100 101

Comentarios a estos manuales

Sin comentarios