Traffic Analyzer Tabs
166 FTB-8500 Series and FTB-8120NGE/8130NGE
Ethernet TX
Rate: Press the Rate field to select the rate for the automated error.
Choices are: 1.0E-02, 1.0E-03, 1.0E-04, 1.0E-05, 1.0E-06, 1.0E-07,
1.0E-08, 1.0E-09 or user definable from 1.0E-09 to 1.0E-02. The default
setting is 1.0E-04.
Continuous: Generates the selected error for each generated frame
when the Continuous check box is selected while the On/Off button is
enabled (On). The Continuous check box is cleared by default.
On/Off button: The On/Off button is used to activate/deactivate the
selected automated error at the rate specified or continuously. This
setting is disabled (Off) by default.