EtherSAM Tabs
292 FTB-8500 Series and FTB-8120NGE/8130NGE
Ramp (Configuration)
Step List
Step indicates the step number. Up to 10 steps may be configured; up
to 7 pre-CIR, CIR, CIR+EIR, and Traffic Policing. Steps are
automatically sorted by the % of CIR value.
Value (% of CIR) indicates a percentage of CIR, CIR+EIR, and/or
Traffic Policing RX rate (throughput) level. CIR+EIR is part of the step
list as soon as it is enabled for at least one service. However, the
CIR+EIR step will only be part of the test for services that have the
CIR+EIR check box selected.
Edit button allows to edit a ramp step. To modify a ramp step, press
the Edit button and enter the new CIR percentage.
Add button allows to add a new ramp step. To add a new ramp step,
press he Add button and enter the CIR percentage.
Delete button allows to delete the selected ramp step. To delete a
ramp step, select the ramp step from the list and press on the Delete
button. CIR, CIR+EIR and Traffic Policing steps cannot be deleted. To
remove CIR+EIR, CIR, and/or Traffic Policing from the list, clear the
corresponding check box from the Services tab for all services.
Default button allows to return to the default ramp steps.
Step Time
The step time represents the test duration for each ramp step (from 5 to 60
seconds).
Ramp Duration
The ramp duration indicates the total time required to perform all the ramp
steps for each service.