Configuring the FTB-7000D OTDR on the IQS-12002B
Calibration System 87
Creating the GP-121723: 50-m Launch Fiber as a Calibration Instrument
1c. Enter the inventory number and comments if necessary.
Note: When the basic model of the calibration instrument is configured to appear
in the Calibration Certificate produced by the IQS-12002B, the information
that you enter in the Inventory number text box will appear in the
Calibration Certificate in the section entitled Standards used to establish
traceability.
1d. Select Calibrated (no recalibration required).
2. Click OK.
3. From the Add a Calibration Event window, enter the manufacturing
date (if the day of manufacturing is not available, enter the first day of
the manufacturing month), and click OK.