Configuring the FTB-7000D OTDR on the IQS-12002B
98 IQS-12002B
Creating the GP-121310: Calibration Instrument for Singlemode, Front-Connector Position
Creating the GP-121310: Calibration
Instrument for Singlemode, Front-Connector
Position (APC Connectors)
This section presents the procedure to create the calibration instrument for
the singlemode, front-connector position for APC connectors.
To create the calibration instrument:
1. From the Configuration menu, select Calibration Instruments >
Standards and click Add.
2. From the Add a standard window, enter the relevant information:
2a. Match the calibration instrument appearing in the Calibration
Instrument list to the instrument you intend to create.
The calibration name will be displayed in the Calibration
Instrument list.
2b. Enter the instrument serial number followed by “(APC)”
(e.g., 123456 (APC)).
2c. Enter the inventory number and comments if necessary.