Exfo IQS-12002B Calibration System Manual de usuario Pagina 61

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 108
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 60
Calibration System 57
A Calibration Methods
This chapter presents the methods, on which are based the procedures of
the IQS-12002B Calibration System., for the calibration of power meters
and OTDRs.
Power Meter
The calibration of a power meter includes the calibration of its absolute
power and, for a high-end unit (e.g., IQS power meter), the calibration of
its linearity.
Absolute Power Calibration
Power meter calibration is performed by comparing it with an
IQS-1502 Calibration Power Meter (working standard). Comparison
between the two is done by measuring the output power of a laser source,
such as the IQS-2400 (usually a DFB laser), having excellent wavelength
and power stability.
In order to improve accuracy, the system can integrate a polarization
scrambler, which reduces the uncertainty due to the polarization
dependant response (PDR) of photo detectors.
For calibration at the 850-nm wavelength, the IQS-2400 uses a stabilized FP
laser with the mandatory mode scrambler in order to obtain a stable laser
source.
Vista de pagina 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 107 108

Comentarios a estos manuales

Sin comentarios