Setting Up Your Calibration System
22 IQS-12002B
Configuring the Due-for-Calibration Reminder
Configuring the Due-for-Calibration Reminder
Since the system relies on the accuracy of standards, the application will
warn you before a standard is due for recalibration (by default, 2 months
ahead). You can set the number of months before the application reminds
you that a recalibration will be necessary. For information on how to
proceed when a standard is back from calibration, see Using a Standard
after Recalibration on page 54.
To configure the due-for-calibration reminder:
1. From the Tools menu, select System > Configure
Due-for-Calibration Reminder.
2. Enter the desired number of months and click OK.