User GuideLight Source for IQS PlatformsIQS-2100
Light Source 52 Safety InformationYour instrument is a Class 1 laser or LED product in compliance with standards IEC 60825-1 Amendment 2: 2001 and 21
Light Source 73 Getting Started with Your Light SourceInserting and Removing Test ModulesTo insert a module into the controller or expansion unit:1. E
Getting Started with Your Light Source8 IQS-2100Inserting and Removing Test Modules4. Insert the protruding edges of the module into the grooves of th
Getting Started with Your Light SourceLight Source 9Inserting and Removing Test ModulesThe module is correctly inserted when its front panel is flush
Getting Started with Your Light Source10 IQS-2100Inserting and Removing Test Modules3. Cover empty slots with the supplied protective covers.3a. Slide
Getting Started with Your Light SourceLight Source 11Starting the Light Source ApplicationStarting the Light Source ApplicationYour IQS-2100 Light Sou
Getting Started with Your Light Source12 IQS-2100Starting the Light Source ApplicationNote: Pressing the LED push button will not activate or turn on
Getting Started with Your Light SourceLight Source 13Starting the Light Source ApplicationTitle BarThe title bar is located at the top of the main win
ii IQS-2100Copyright © 2003–2008 EXFO Electro-Optical Engineering Inc. All rights reserved. No part of this publication may be reproduced, stored in a
Getting Started with Your Light Source14 IQS-2100Starting the Light Source ApplicationData DisplayIn addition to the Control Center, the main window a
Getting Started with Your Light SourceLight Source 15Entering Values Using Sliders and Numeric BoxesEntering Values Using Sliders and Numeric BoxesMan
Getting Started with Your Light Source16 IQS-2100Exiting the ApplicationExiting the ApplicationClosing any application that is not currently being use
Light Source 174 Setting Up Your Light SourceLight source features are controlled with the Windows-compatible IQS Manager software. Please refer to th
Setting Up Your Light Source18 IQS-2100Selecting a WavelengthSelecting a WavelengthIf you are using a dual-wavelength module, you can select a wavelen
Setting Up Your Light SourceLight Source 19Selecting a WavelengthTo select the desired wavelength:Click the appropriate button in the Wavelength panel
Setting Up Your Light Source20 IQS-2100Setting the AttenuationSetting the AttenuationYou can modify the power of the IQS-2100 Light Source output by a
Setting Up Your Light SourceLight Source 21Setting the AttenuationThe attenuation value in the data display will then indicate the increased or decrea
Setting Up Your Light Source22 IQS-2100Selecting a Modulation FrequencySelecting a Modulation FrequencyYou can modulate the laser output of the IQS-21
Setting Up Your Light SourceLight Source 23Saving and Recalling ConfigurationsSaving and Recalling ConfigurationsOnce you have set the IQS-2100 Light
ContentsLight Source iiiContentsCertification Information ...
Setting Up Your Light Source24 IQS-2100Saving and Recalling ConfigurationsTo recall a configuration:1. Select the Configuration function tab.2. Click
Light Source 255 Operating Your Light SourceOnce you have set your source parameters, you are ready to use the source in a test setup.To operate the s
Operating Your Light Source26 IQS-2100Installing the EXFO Universal Interface (EUI)If your connector features a screwsleeve, tighten the connector eno
Operating Your Light SourceLight Source 27Activating or Deactivating Light EmissionActivating or Deactivating Light EmissionBefore activating the sour
Operating Your Light Source28 IQS-2100Viewing ResultsNote: You should let the source warm up for 30 minutes to obtain optimum wavelength stability.You
Light Source 296 Monitoring Light Source ModulesWhen using your IQS-2100 Light Source module, either alone or with other modules in a test setup, you
Monitoring Light Source Modules30 IQS-2100Using Monitor WindowsTo select modules and display their monitor windows:1. On the Current Modules function
Monitoring Light Source ModulesLight Source 31Using Monitor WindowsRearrangeMonitorsbutton(1 or 2columns)Close All buttonRemoveMonitorbuttonMonitor wi
Monitoring Light Source Modules32 IQS-2100Using QuickToolsUsing QuickToolsWith QuickTools, you can fine-tune your module directly, while keeping an ey
Monitoring Light Source ModulesLight Source 33Using QuickToolsFor the IQS-2100 Light Source, one of the two QuickTools utility versions will be displa
Contentsiv IQS-21008 Troubleshooting ...43Solving Common Proble
Monitoring Light Source Modules34 IQS-2100Using QuickTools³ From the Modulation section, select a modulation frequency by clicking the selection arrow
Light Source 357 MaintenanceTo help ensure long, trouble-free operation:³ Always clean fiber-optic connectors before using them.³ Keep the unit free o
Maintenance36 IQS-2100Cleaning Fixed ConnectorsCleaning Fixed ConnectorsRegular cleaning of connectors will help maintain optimum performance. Do not
MaintenanceLight Source 37Cleaning Fixed Connectors7. Slowly insert the cleaning tip into the connector until it reaches the ferrule inside (a slow cl
Maintenance38 IQS-2100Cleaning EUI ConnectorsCleaning EUI ConnectorsRegular cleaning of EUI connectors will help maintain optimum performance. There i
MaintenanceLight Source 39Cleaning EUI Connectors5. Repeat steps 3 to 4 with a dry cleaning tip.Note: Make sure you don’t touch the soft end of the cl
Maintenance40 IQS-2100Recalibrating the UnitRecalibrating the UnitManufacturing and service center calibrations are based on the ISO/IEC 17025 Standar
MaintenanceLight Source 41Recycling and Disposal (Applies to European Union Only)Recycling and Disposal (Applies to European Union Only)³ Unless other
Light Source 438 TroubleshootingSolving Common ProblemsThe following is a list of common problems along with their possible causes and some recommende
Certification InformationLight Source vCertification InformationF.C.C. InformationElectronic test equipment is exempt from Part 15 compliance (FCC) in
Troubleshooting44 IQS-2100Viewing Online DocumentationViewing Online DocumentationAn online version of the IQS-2100 Light Source user guide is availab
TroubleshootingLight Source 45Contacting the Technical Support GroupContacting the Technical Support GroupTo obtain after-sales service or technical s
Troubleshooting46 IQS-2100Contacting the Technical Support Group³ Select the Technical Support tab to view phone numbers and active Internet links to
TroubleshootingLight Source 47TransportationTransportationMaintain a temperature range within specifications when transporting the unit. Transportatio
Light Source 499 WarrantyGeneral InformationEXFO Electro-Optical Engineering Inc. (EXFO) warrants this equipment against defects in material and workm
Warranty50 IQS-2100ExclusionsExclusionsEXFO reserves the right to make changes in the design or construction of any of its products at any time withou
WarrantyLight Source 51Service and RepairsService and RepairsEXFO commits to providing product service and repair for five years following the date of
Warranty52 IQS-2100EXFO Service Centers WorldwideEXFO Service Centers WorldwideIf your product requires servicing, contact your nearest authorized ser
Light Source 53A Technical SpecificationsIMPORTANTThe following technical specifications can change without notice. The information presented in this
Certification Informationvi IQS-2100Application of Council Directive(s): 73/23/EEC - The Low Voltage Directive89/336/EEC - The EMC DirectiveAnd their
Light Source 55B SCPI Command ReferenceThis appendix presents detailed information on the commands and queries supplied with your IQS-2100 Light Sourc
SCPI Command Reference56 IQS-2100Quick Reference Command TreeQuick Reference Command TreeCommand Parameter(s) P.SNUMber? 57SOURce[1..n] AM INTernal FR
SCPI Command ReferenceLight Source 57Product-Specific Commands—DescriptionProduct-Specific Commands—Description:SNUMber?Description This query returns
SCPI Command Reference58 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:AM:INTernal:FREQuencyDescription This command selects the internal
SCPI Command ReferenceLight Source 59Product-Specific Commands—Description:SOURce[1..n]:AM:INTernal:FREQuency?Description This query returns a value i
SCPI Command Reference60 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:POWer:ATTenuationDescription This command changes the internal att
SCPI Command ReferenceLight Source 61Product-Specific Commands—DescriptionExample(s) SOUR:POW:STAT ON SOUR:POW:ATT 2 Notes Instrument must be in the R
SCPI Command Reference62 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:POWer:ATTenuation?Description This query returns a value correspon
SCPI Command ReferenceLight Source 63Product-Specific Commands—DescriptionResponse(s) Attenuation: The response data syntax for <Attenuation> is
Light Source 11 Introducing the IQS-2100 Light SourceMain FeaturesThe IQS-2100 Light Source is designed for scientific and industrial applications usi
SCPI Command Reference64 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:POWer:STATeDescription This command turns the optical source on or
SCPI Command ReferenceLight Source 65Product-Specific Commands—Description:SOURce[1..n]:POWer:STATe?Description This query returns a value indicating
SCPI Command Reference66 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:POWer:WAVelengthDescription This command selects a wavelength when
SCPI Command ReferenceLight Source 67Product-Specific Commands—Description:SOURce[1..n]:POWer:WAVelength?Description This query returns a value indica
SCPI Command Reference68 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:POWer:WAVelength:COUNt?Description This query returns the number o
SCPI Command ReferenceLight Source 69Product-Specific Commands—Description:SOURce[1..n]:POWer:WAVelength:LOWer?Description This query returns a value
SCPI Command Reference70 IQS-2100Product-Specific Commands—Description:SOURce[1..n]:POWer:WAVelength:UPPer?Description This query returns a value indi
SCPI Command ReferenceLight Source 71Product-Specific Commands—Description:STATus?Description This query returns a value indicating the status of the
IndexLight Source 73IndexAAbout function tab ... 45activating the light source... 27active L
Introducing the IQS-2100 Light Source2 IQS-2100Available ModelsA single-operation display screen lets you activate sources and select signal wavelengt
Index74 IQS-2100Llabel, identification ... 45LabVIEW drivers ... 2LED push
IndexLight Source 75wavelength, light source ... 18serial number, module ... 45service and repairs...
www.EXFO.com · [email protected]/N: 1052814CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADATel.: 1 418 683-0211 · Fax: 1 418 683-
Introducing the IQS-2100 Light SourceLight Source 3ConventionsConventionsBefore using the product described in this manual, you should understand the