Exfo FTB-7000 OTDR Series for FTB-500 Manual de usuario Pagina 419

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 482
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 418
SCPI Command Reference
OTDR 405
Product-Specific Commands—Description
:FETCh[1..n]:RANGe?
Description This query returns the range for the trace
corresponding to the specified trace index.
*RST clears this setting.
Syntax :FETCh[1..n]:RANGe?<wsp>TRC1|TRC2|TRC3|
TRC4
Parameter(s) Label:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
TRC1|TRC2|TRC3|TRC4.
Trace index of the available wavelengths.
Response Syntax <Range>
Response(s) Range:
The response data syntax for <Range> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the range, in meters.
Example(s) CONF:ACQ 1310,NM1250,M10 NS
CONF:ACQ:MODE ACQ
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
FETC:RANG? TRC1 Returns 1.25E+3
See Also FETCh[1..n]:ASETting:RANGe?
MMEMory[1..n]:LOAD:TRACe
TRACe[1..n]:CATalog?
Vista de pagina 418
1 2 ... 414 415 416 417 418 419 420 421 422 423 424 ... 481 482

Comentarios a estos manuales

Sin comentarios