Exfo FTB-7000 OTDR Series for FTB-500 Manual de usuario Pagina 406

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 482
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 405
SCPI Command Reference
392 OTDR
Product-Specific Commands—Description
:CONFigure[1..n]:ANAlysis:THReshold:
REFLectance?
Description This query returns the reflectance threshold that
will be used for the next acquisition.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ANAlysis:THReshold:REFLecta
nce?[<wsp>MINimum|MAXimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Reflectance>
Response(s) Reflectance:
The response data syntax for <Reflectance> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the reflectance threshold.
Example(s) CONF:ANA:THR:REFL -72.5
CONF:ANA:THR:REFL? Returns -72.5
Vista de pagina 405
1 2 ... 401 402 403 404 405 406 407 408 409 410 411 ... 481 482

Comentarios a estos manuales

Sin comentarios