Exfo FTB-7000 OTDR Series for FTB-500 Manual de usuario Pagina 364

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 482
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 363
SCPI Command Reference
350 OTDR
Product-Specific Commands—Description
Parameter(s) ³ MarkerB:
The program data syntax for <MarkerB> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Specifies the marker B position, in meters.
³ SubMarkerB:
The program data syntax for <SubMarkerB> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Specifies the submarker B position, in meters.
Return the splice loss value, calculated using all
four markers.
Response Syntax <Splice Loss>
Response(s) Splice Loss:
The response data syntax for <Splice Loss> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Return the splice loss value, calculated using all
four markers.
Example(s) CONF:ACQ:MODE ACQUISITION
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
CALC:SLOS? TRC1,10,100,200,300 Ex.: Returns
0.058
CALC:SLOS? TRC1,0.01 KM,100 M, 0.2 KM,300 Ex.:
Returns 0.058
:CALCulate[1..n]:SLOSs?
Vista de pagina 363
1 2 ... 359 360 361 362 363 364 365 366 367 368 369 ... 481 482

Comentarios a estos manuales

Sin comentarios