Exfo FTB-7000 OTDR Series for FTB-500 Manual de usuario Pagina 411

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 482
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 410
SCPI Command Reference
OTDR 397
Product-Specific Commands—Description
:FETCh[1..n]:ASETting:DURation?
Description This query returns the duration found after an
initiate (INIT) command. Note that acquisition
mode (CONF:ACQ:MODE) must be set to
ASETting.
Since *RST clears the duration value, the
returned value will be 0.
Syntax :FETCh[1..n]:ASETting:DURation?
Parameter(s) None
Response Syntax <Duration>
Response(s) Duration:
The response data syntax for <Duration> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
Returns the duration, in seconds.
Example(s) CONF:ACQ:MODE ASET
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
FETC:ASET:DUR? Ex.: Returns 15
See Also CONFigure[1..n]:ACQuisition:DURation?
Vista de pagina 410