SCPI Command Reference
OTDR 339
Product-Specific Commands—Description
Response(s) Loss:
The response data syntax for <Loss> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the loss value in dB, between marker A
and marker B.
Example(s) CONF:ACQ:MODE ACQUISITION
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
CALC:LOSS? TRC1,10,104 Ex.: Returns 0.458
CALC:LOSS? TRC1,10 M,0.104 KM Ex.: Returns
0.458
CALC:LOSS? TRC1,0.01 KM,104 M Ex.: Returns
0.458
See Also MMEMory[1..n]:LOAD:TRACe
TRACe[1..n]:CATalog?
:CALCulate[1..n]:LOSS?